QWhat is the smallest defect size that can be seen in the CT data scanned by the FXO-CT2000 device?

AGenerally speaking, the minimum defect size that can be identified is greater than the spatial resolution value. According to practical experience, it is basically 2 times or even more than 3 times the pixel size to identify. Since the individual pixel size of the data and the spatial resolution size are equal, if the customer wants to see a sample with a minimum defect of 100 microns, we must control the spatial resolution (accuracy) between 1-50 microns when we perform a CT scan, the higher the resolution, the more visible and the clearer the defect.

QFXO-CT2000 Device What is the maximum accuracy (spatial resolution) of the device?

ABecause the focus size of the 225kv micron tube is 5 microns, the JIMA card can only see 5 microns at present, so the theoretical maximum accuracy is to 5 microns.

Q工业CT设备的组成结构是什么?

A射线源、机械扫描系统、探测器系统、计算机系统、屏蔽设施等组成。

Q工业CT设备的工作原理是什么?

A在检测过程中,射线管发射x射线,同时载物台上的样品会进行一定角度的旋转,探测器接收衰减的x射线信号并转化成电信号,数据采集系统再把电信号转化成数字信号。最后通过计算机系统对数字信号进行处理和计算,生成被检物体的CT内部图像。

QDR和CT的区别是什么?

ADR是通过单次曝光生成的一张平面图,图像会把所有信息压缩在同一层上,导致个别缺陷因为角度的问题或前后的叠加无法看见,而且分辨率低。而CT是多次多角度的曝光并通过软件生成三维图像,可以看任意方向的切面,所有你需要的信息都能看见。

QX射线的穿透力数据是多少?

AX射线的穿透深度和样品的材料密度、质量吸收系数有关系。理论上225kv的反射式靶射线管可以穿透5cm的钢和20cm的铝。