CN
ABOUT
Profile
Culture
Patent
CT & DR
Precise Desktop CT
Versatile Cabinet CT
Versatile Precise CT
Robust Precise CT
Dual X-ray Tube CT
Megavolt Industrial CT
Micro-focal Spot AXI DR
SECURITY SOLUTIONS
FXO-100100 Inspetion System
FXO-150180 Inspection System
FXO-180180 Inspection System
FXO-BMS Vehicle Inspection System
FXO-HCBS Handheld Scanner
Stationary Vehicle Inspection System
APPLICATIONS
Aerospace
Defense & Shipbuild
3D Printing
Automobile
IC & Electronics
New Materials
Lithium Battery
Laboratories
Public Safety
Geology
Archaeology & Museology
Life Science
SERVICES & SUPPORT
Testing Services
Technical Support
How We Service
FAQ
NEWS
CONTACT
Contact
Job Openings
Testing Services
Technical Support
How We Service
FAQ
Your current location:
Home
>
SERVICES & SUPPORT
>
FAQ
Q
What is the smallest defect size that can be seen in the CT data scanned by the FXO-CT2000 device?
A
Generally speaking, the minimum defect size that can be identified is greater than the spatial resolution value. According to practical experience, it is basically 2 times or even more than 3 times the pixel size to identify. Since the individual pixel size of the data and the spatial resolution size are equal, if the customer wants to see a sample with a minimum defect of 100 microns, we must control the spatial resolution (accuracy) between 1-50 microns when we perform a CT scan, the higher the resolution, the more visible and the clearer the defect.
Q
FXO-CT2000 Device What is the maximum accuracy (spatial resolution) of the device?
A
Because the focus size of the 225kv micron tube is 5 microns, the JIMA card can only see 5 microns at present, so the theoretical maximum accuracy is to 5 microns.
Q
FXO-CT2000设备设备能扫最大样品的尺寸是多少?
A
520*650mm
Q
FXO-CT2000设备设备能扫样品的重量极限是多少?
A
20KG
Q
工业CT设备的组成结构是什么?
A
射线源、机械扫描系统、探测器系统、计算机系统、屏蔽设施等组成。
Q
工业CT设备的工作原理是什么?
A
在检测过程中,射线管发射x射线,同时载物台上的样品会进行一定角度的旋转,探测器接收衰减的x射线信号并转化成电信号,数据采集系统再把电信号转化成数字信号。最后通过计算机系统对数字信号进行处理和计算,生成被检物体的CT内部图像。
Q
DR和CT的区别是什么?
A
DR是通过单次曝光生成的一张平面图,图像会把所有信息压缩在同一层上,导致个别缺陷因为角度的问题或前后的叠加无法看见,而且分辨率低。而CT是多次多角度的曝光并通过软件生成三维图像,可以看任意方向的切面,所有你需要的信息都能看见。
Q
X射线的穿透力数据是多少?
A
X射线的穿透深度和样品的材料密度、质量吸收系数有关系。理论上225kv的反射式靶射线管可以穿透5cm的钢和20cm的铝。
Prev
1
Next
service@focus-sh.net
(086) 021-31019175
Room 102, Building 1, No.18, Handu Rd., Pudong New Area, Shanghai
ABOUT
CT & DR
SECURITY SOLUTIONS
APPLICATIONS
SERVICES & SUPPORT
NEWS
CONTACT
Copyright © 2023 上海福柯斯智能科技有限公司 All Rights Reserved.
沪ICP备2023025604号-1
Get in touch with an expert
8619729710809
Scan and follow us
Tel:8602131019175